Single X-ray absorptiometry of the forearm: precision, correlation, and reference data

Calcif Tissue Int. 1994 Mar;54(3):212-8. doi: 10.1007/BF00301681.

Abstract

The performance of a single X-ray absorptiometry (SXA) device incorporating an X-ray tube as a photon source was evaluated with respect to precision in vivo and in vitro, scan time, image quality, and correlation with an existing dual energy X-ray absorptiometry (DXA) device. SXA precision in vivo, expressed as a coefficient of variation (CV), was 0.66% for bone mineral content (BMC) and 1.05% for bone mineral density (BMD). Precision in vitro, based on 78 BMC measurements of a forearm phantom over 195 days, was 0.53%. Correlation with DXA at the 8 mm distal forearm site was high (r = 0.97 for BMC and r = 0.96 for BMD). A preliminary SXA reference database composed of 151 healthy Caucasian American women was developed to facilitate the interpretation of patient measurements. SXA scan time was 4 minutes and delivered a radiation exposure of 1.68 mrem. SXA image quality and spatial resolution were superior to SPA and comparable to DXA.

MeSH terms

  • Absorptiometry, Photon*
  • Adult
  • Aged
  • Aged, 80 and over
  • Bone Density*
  • Computer Simulation
  • Female
  • Forearm
  • Humans
  • Middle Aged
  • Osteoporosis / physiopathology
  • Reference Values
  • Reproducibility of Results