Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies

BMJ 2011; 342 doi: https://doi.org/10.1136/bmj.d1855 (Published 14 April 2011) Cite this as: BMJ 2011;342:d1855
  1. Roman Hovorka, principal research associate12,
  2. Kavita Kumareswaran, clinical research fellow 13,
  3. Julie Harris, research nurse1,
  4. Janet M Allen, research nurse12,
  5. Daniela Elleri, clinical research fellow12,
  6. Dongyuan Xing, biostatistician4,
  7. Craig Kollman, senior biostatistician4,
  8. Marianna Nodale, research assistant1,
  9. Helen R Murphy, senior research associate1,
  10. David B Dunger, professor of paediatrics12,
  11. Stephanie A Amiel, professor of diabetic medicine5,
  12. Simon R Heller, professor of diabetes6,
  13. Malgorzata E Wilinska, research associate12,
  14. Mark L Evans, senior lecturer13
  1. 1Institute of Metabolic Science, University of Cambridge, UK
  2. 2Department of Paediatrics, University of Cambridge, UK
  3. 3Department of Medicine University of Cambridge, UK
  4. 4The Jaeb Center for Health Research, Tampa, FL 33647, USA
  5. 5Diabetes Research, Weston Education Centre, King’s College London, London, UK
  6. 6Diabetes Centre, Clinical Sciences Centre, Northern General Hospital, Sheffield, UK
  1. Correspondence to: R Hovorka, Institute of Metabolic Science, University of Cambridge, Box 289, Addenbrooke’s Hospital, Cambridge CB2 0QQ, UK rh347{at}cam.ac.uk
  • Accepted 4 February 2011

Abstract

Objective To compare the safety and efficacy of overnight closed loop delivery of insulin (artificial pancreas) with conventional insulin pump therapy in adults with type 1 diabetes.

Design Two sequential, open label, randomised controlled crossover, single centre studies.

Setting Clinical research facility.

Participants 24 adults (10 men, 14 women) with type 1 diabetes, aged 18-65, who had used insulin pump therapy for at least three months: 12 were tested after consuming a medium sized meal and the other 12 after consuming a larger meal accompanied by alcohol.

Intervention During overnight closed loop delivery, sensor measurements of glucose were fed into a computer algorithm, which advised on insulin pump infusion rates at 15 minute intervals. During control nights, conventional insulin pump settings were applied. One study compared closed loop delivery of insulin with conventional pump therapy after a medium sized evening meal (60 g of carbohydrates) at 1900, depicting the scenario of “eating in.” The other study was carried out after a later large evening meal (100 g of carbohydrates) at 2030, accompanied by white wine (0.75 g/kg ethanol) and depicted the scenario of “eating out.”

Main outcome measures The primary outcome was the time plasma glucose levels were in target (3.91-8.0 mmol/L) during closed loop delivery and a comparable control period. Secondary outcomes included pooled data analysis and time plasma glucose levels were below target (≤3.9 mmol/L).

Results For the eating in scenario, overnight closed loop delivery of insulin increased the time plasma glucose levels were in target by a median 15% (interquartile range 3-35%), P=0.002. For the eating out scenario, closed loop delivery increased the time plasma glucose levels were in target by a median 28% (2-39%), P=0.01. Analysis of pooled data showed that the overall time plasma glucose was in target increased by a median 22% (3-37%) with closed loop delivery (P<0.001). Closed loop delivery reduced overnight time spent hypoglycaemic (plasma glucose ≤3.9 mmol/L) by a median 3% (0-20%), P=0.04, and eliminated plasma glucose concentrations below 3.0 mmol/L after midnight.

Conclusion These two small crossover trials suggest that closed loop delivery of insulin may improve overnight control of glucose levels and reduce the risk of nocturnal hypoglycaemia in adults with type 1 diabetes.

Trial registration ClinicalTrials.gov NCT00910767 and NCT00944619.

Introduction

Worldwide, type 1 diabetes accounts for 5-15%1 of 285 million adults with diabetes.2 The incidence of type 1 diabetes is increasing at a rate of 3% per year, with the highest incidence in white European populations, particularly those in northern Europe.3 Lifelong insulin replacement is required, but conventional regimens for delivering insulin fail to achieve physiological glucose levels resulting in an increased risk of long term vascular complications.4 Although educational interventions have a modest effect on outcomes these are difficult to sustain and cannot alone facilitate optimal glucose control.5 A major challenge of intensive insulin therapy is the risk of hypoglycaemia,6 with low blood glucose levels typically lasting more than two hours a day,7 impeding efforts to optimise glucose control and reducing quality of life.6

Since early 2000, improved tools for managing type 1 diabetes have emerged. These include continuous glucose monitors and increasingly sophisticated insulin pumps. Continuous glucose monitors measure interstitial glucose levels and, although less accurate than conventional monitoring by finger stick testing, they offer unprecedented insights into daily glucose excursions8 and their regular use improves glucose control.9 Insulin pumps are designed to mimic physiological delivery of insulin, with pre-meal doses determined according to the carbohydrate content of a planned meal, glucose levels from finger stick testing, and insulin sensitivity factors.10 Basal doses are delivered at pre-programmed rates over a 24 hour period and are intermittently adjusted according to finger stick glucose levels. Combining continuous glucose monitoring with pump therapy can improve the control of glucose,11 but responding to overnight real time data detected by sensor is impractical. More than half of hypoglycaemic events occur overnight12 and the risk of hypoglycaemia increases with alcohol ingestion.13 Thus developing measures to prevent nocturnal hypoglycaemia is a priority.

Recent technological advances provide an opportunity to combine three technologies—continuous glucose monitors, insulin pumps, and a computerised control algorithm. This forms the basis of a closed loop system, sometimes termed an artificial pancreas, and characterised by algorithm driven insulin delivery according to glucose levels detected by a sensor.14 15 16 Research is under way to develop closed loop systems of increasing complexity.17 We have already shown that overnight closed loop delivery of insulin can improve glucose control and reduce the risk of nocturnal hypoglycaemia in children and adolescents.18 Here, we hypothesise that these findings can be extended to adults.

Methods

We carried out two sequential, open label, randomised controlled crossover studies comparing the overnight closed loop delivery of insulin with conventional insulin pump (continuous subcutaneous insulin infusion) therapy (fig 1). The studies mimicked the two common scenarios of “eating in” and “eating out” in preparation for testing as an outpatient. In an initial study we evaluated closed loop delivery after a medium sized evening meal, mimicking an evening at home when closed loop delivery could be conveniently started with the meal. In a second, more challenging, study mimicking an evening out, the meal was larger, was consumed later, and was accompanied by wine. Closed loop delivery was started later and continued until noon the next day.

Figure1

Fig 1 Design of two studies comparing closed loop delivery of insulin with conventional insulin pump therapy after two meal scenarios: eating in and eating out

Setting and participants

The studies were carried out at the Wellcome Trust clinical research facility at Addenbrooke’s Hospital, Cambridge, between February and December 2009. Inclusion criteria were adults with type 1 diabetes (World Health Organization criteria or confirmed negative for C peptide), aged 18-65, and using insulin pump therapy for at least three months. Exclusion criteria were awareness of reduced hypoglycaemia and clinically significant nephropathy, neuropathy, or retinopathy. For the eating out scenario, we recruited participants who could tolerate the amount of alcohol used and excluded participants with insulin resistance (total daily insulin dose ≥1.4 U/kg), with poor glycaemic control (HbA1c ≥10%), and who were pregnant or breast feeding. Participants signed informed consent. Figure 2 shows the flow of participants through the study.

Figure2

Fig 2 Flow of participants through study comparing closed loop delivery of insulin with conventional insulin pump therapy after a medium sized evening meal (eating in scenario), and study comparing closed loop delivery with insulin pump therapy after a large evening meal accompanied by alcohol (eating out scenario)

Eating in scenario

Of 32 adults invited to take part in the study, 13 agreed and were randomly assigned to be treated overnight with either closed loop delivery of insulin or conventional insulin pump therapy during two study nights, separated by an interval of one to three weeks. Before the first study visit we analysed data from continuous glucose monitoring for up to 120 hours to optimise the delivery of insulin by conventional pump therapy. On both visits participants ate an identical evening meal, comprising 60 g of carbohydrate, at 1900, accompanied by prandial insulin. We calculated the prandial boluses according to the participants’ own insulin to carbohydrate ratio and glucose values from finger stick testing. During the intervention visit, closed loop insulin delivery was applied from 1900 until 0800 the next day. During the control visit, participants applied their usual insulin pump settings over the same timeframe.

Eating out scenario

We approached a further 41 adults with type 1 diabetes and recruited 12 to a second randomised crossover study of closed loop delivery or conventional insulin pump therapy after a large evening meal accompanied by alcohol, depicting the scenario of “eating out.” On both visits, one to three weeks apart, participants ate an identical evening meal, comprising 100 g of carbohydrate, at 2030, accompanied by prandial insulin and dry white wine (Chenin Blanc, 13% vol, Ken Forrester Wines, South Africa). Participants drank the wine (7.2 mL/kg or 6.6 units per 70 kg participant) with or after their meal, completing the meal by 2200. During the intervention visit, insulin was delivered by closed loop from 2200 until 1200 the next day. During the control visit, participants applied their usual insulin pump settings over the same timeframe.

Continuous glucose monitoring and insulin delivery

To measure subcutaneous glucose in the eating in scenario we used the continuous glucose monitoring system FreeStyle Navigator (Abbott Diabetes Care, Alameda, CA) with a 10 hour run-in calibration period. In the eating out scenario we used FreeStyle Navigator with a one hour run-in calibration period. The systems were calibrated using capillary finger stick measurements as per manufacturer’s instructions. The accuracy of the sensor, evaluated as the median relative absolute difference between sensor glucose levels and paired plasma glucose levels divided by plasma glucose levels, was 8.0% (4.5-19.3%) in the eating in scenario and 12.0% (6.8-17.2%) in the eating out scenario.

When participants arrived at the research clinic, we replaced their insulin pump with a study pump (Deltec Cozmo; Smiths Medical, St Paul, MN). This pump was connected to the established infusion site, delivering rapid acting insulin analogue Aspart (Novo Nordisk, Bagsvaerd, Denmark).

Closed loop algorithm

We used an algorithm based on the model predictive control approach.18 19 Every 15 minutes a research nurse initiated a control cycle; the nurse inputted the sensor glucose value into the computer based algorithm and adjusted the insulin pump according to the basal infusion rate calculated by the algorithm. The calculations utilised a compartment model of glucose kinetics,20 describing the effect of rapid acting insulin and the carbohydrate content of meals on glucose excursions detected by the sensor. The algorithm was initialised using participant’s weight, total daily insulin dose, and basal insulin requirements. Additionally, the algorithm was provided with glucose levels measured by the sensor during a 30 minute period preceding the start of closed loop delivery, the carbohydrate content of the evening meal, and the prandial insulin bolus. The algorithm adapted itself to participants by updating two model variables: an endogenous glucose flux correcting for errors in model based predictions, and carbohydrate bioavailability. Several competing models differing in the absorption of subcutaneous insulin and oral carbohydrates ran in parallel.21 A combined model forecasted plasma glucose excursions over a 2.5 hour prediction horizon. The algorithm aimed to achieve glucose levels between 5.8 and 7.3 mmol/L and adjusted the actual level depending on fasting versus postprandial status, preceding glucose levels, and the accuracy of predictions made by the compartment model. Safety rules limited maximum insulin infusion and suspended insulin delivery when the sensor measured glucose at or below 4.3 mmol/L or when the sensor detected that glucose was decreasing rapidly. We used algorithm version 0.02.04 to 0.02.18.

Plasma glucose and plasma insulin sampling

We collected venous samples for glucose determination every 15 minutes and for insulin assay every 30 minutes. In the eating in scenario we collected samples between 1830 and 0800 and in the eating out scenario between 1930 and 1200. These data were not used to alter insulin infusion rates during the visits for closed loop delivery or control. In the eating out scenario we collected additional samples for determination of ethanol at 2030, 2200, and midnight and then three hourly until 1200 the next day.

Assays

We used YSI2300 STAT Plus Analyser (YSI, Lynchford House, Farnborough, UK) to determine plasma glucose levels, and an immunochemiluminometric assay (Invitron; Monmouth, UK) with an intra-assay coefficient of variation of 4.7% and an interassay coefficient of variation of 7.2% to 8.1% for plasma insulin assay. We determined plasma ethanol levels using the ethyl alcohol method (Dade Behring, Atterbury, UK), with an intra-assay coefficient of variation of 2.4% and an interassay coefficient of variation of 5.71%.

Sample size

Based on previous studies18 we anticipated that overnight closed loop delivery would increase the time that plasma glucose concentrations were between 3.91 and 8.00 mmol/L by a mean 37% (SD 40%). We calculated that 12 participants would provide 80% power at the 5% level of significance to detect this difference between conventional and closed loop insulin delivery systems.

Randomisation and masking

For each study we placed a computer generated allocation sequence with permuted block four randomisation in sealed envelopes. Participants were randomised on recruitment. During the eating in scenario, investigators were blinded to the plasma glucose data. For safety reasons, during the eating out scenario investigators had access to plasma glucose levels. During both studies, participants did not have access to data on plasma or sensor glucose levels.

Statistical analysis

Senior investigators and study statisticians agreed on the analysis plan in advance. The primary outcome was the percentage time plasma glucose concentrations were in the target range (3.91-8.00 mmol/L) between 1900 and 0800 in the eating in scenario and between 2200 and 1200 in the eating out scenario. Secondary outcomes were the mean glucose concentration, time when glucose concentration was 3.9 mmol/L or less (hypoglycaemia), time when glucose concentration was greater than 8.0 mmol/L (hyperglycaemia), mean rate of insulin infusion, and mean plasma insulin concentration for both separate and pooled study data. We estimated glycaemic variability by mean amplitude of glycaemic excursions22 and standard deviation of glucose concentration. Secondary outcomes were calculated for both plasma and sensor glucose levels as well as for the period from midnight to end of closed loop control.

We used the low blood glucose index to assess the duration and extent of hypoglycaemia, calculated as an average of transformed glucose measurements progressively increasing at low glucose levels.23 Safety assessments examined plasma glucose levels below 3 mmol/L and above 16.7 mmol/L.

For each outcome we fit a repeated measures regression model based on the ranked normal transformation (except for the mean glucose concentration which was not transformed because it already had an approximate normal distribution) to compare the two treatments adjusting for plasma glucose level at the start of closed loop delivery. We carried out analyses using SAS software, version 9.1 and SPSS, version 15. Kernel density of plasma glucose was estimated using the package “np” (non-parametric kernel smoothing methods for mixed data types), version 0.40-1, adopting a bandwidth of 0.25 mmol/L and implemented in the R, version 2.11.1 (R Foundation for Statistical Computing).

Results

Eating in scenario

One participant who was unable to attend the second visit for logistical reasons withdrew and was excluded from the analysis. Table 1 shows the baseline characteristics of the remaining participants. Figure 3 shows the profiles of plasma glucose and plasma insulin concentrations and insulin infusion. The time spent with plasma glucose levels in target range increased by 15% (3-35%) during closed loop delivery (P=0.002; table 2). Closed loop delivery decreased glucose variability overnight (reduced standard deviation of plasma glucose 1.2 v 1.8 mmol/L, P=0.02; table 3). It also decreased the time spent hyperglycaemic (11% v 27%, P=0.01) and the low blood glucose index (0.8 v 1.5, P=0.03). Insulin infusion was more variable during closed loop delivery, but the average insulin infusion rate and plasma insulin concentrations were similar during closed loop delivery and control visits (fig 3).

Figure3

Fig 3 Profiles (medians and interquartile ranges) of plasma glucose and insulin concentrations and insulin infusion in eating in scenario (12 participants). Outlying squares represent hypoglycaemic events (glucose level <3.0 mmol/L)

Table 1

 Baseline characteristics of participants completing two meal scenarios to compare closed loop delivery of insulin with conventional insulin pump therapy. Values are means (standard deviations) unless stated otherwise

View this table:
Table 2

 Time plasma glucose level in target during closed loop delivery of insulin and conventional insulin pump therapy in two meal scenarios. Values are medians (interquartile ranges) unless stated otherwise

View this table:
Table 3

 Secondary comparisons in eating in scenario based on plasma glucose concentrations. Values are medians (interquartile ranges) unless stated otherwise

View this table:

From midnight, when participants were sleeping, closed loop delivery achieved near normal glycaemia, with plasma glucose levels in target range for 92% of the time (interquartile range 70-100%) compared with 45% (27-80%) on control nights, with a difference of 29% (6-46%) favouring closed loop delivery (P=0.003).

The eating out scenario

Table 1 shows the characteristics of the participants. Use of alcohol during closed loop delivery and control visits was comparable (peak ethanol concentration 15.0 (13.7-18.0) mmol/L v 16.5 (14.2-20.8) mmol/L) as was time to peak ethanol concentration (150 (90-210) minutes v 90 (90-90) minutes) and area under the concentration curve per 15.5 hours (4151 (3375-4779) mmol/L v 4766 (3020-5374) mmol/L). Alcohol ingestion did not affect the performance of closed loop delivery, which increased the time plasma glucose was in target range by 22% (3-37%), P=0.01.

Table 4 shows secondary outcomes, all favouring closed loop delivery of insulin. As for the eating in scenario, the average overnight infusion of insulin and plasma insulin concentration during closed loop delivery and pump therapy were comparable. The profiles for glucose and insulin levels showed lower variability of plasma glucose during closed loop delivery from 0200 onwards (fig 4).

Figure4

Fig 4 Profiles (medians and interquartile ranges) of plasma glucose and insulin concentrations and insulin infusion in the eating out scenario (12 participants). Outlying squares represent hypoglycaemic events (glucose level <3.0 mmol/L)

Table 4

 Secondary comparisons in eating out scenario based on plasma glucose concentrations. Data are medians (interquartile ranges) unless stated otherwise

View this table:

Hypoglycaemic and hyperglycaemic events

During closed loop delivery, four hypoglycaemic events (glucose level <3.0 mmol/L) were observed, two in the eating in scenario and two in the eating out scenario. All occurred before midnight and all were associated with symptoms. Three were attributable to the preceding prandial insulin dose and could not be prevented despite suspended insulin delivery immediately after the start of closed loop. Three episodes were treated with 15-36 g of oral carbohydrates in drink (Lucozade; GlaxoSmithKline, Brentford, UK) and the fourth lasted 30 minutes with spontaneous recovery. When data collected after midnight, while participants slept, were combined (968 readings), no plasma glucose measurement was below 3.0 mmol/L, 7 (<1%) were below 3.3 mmol/L, and 8 (<1%) were below 3.5 mmol/L.

During insulin pump therapy, one hypoglycaemic episode with glucose level below 2.0 mmol/L occurred at 0215 in the eating out scenario. The episode was not associated with symptoms but was treated with 30 g of oral carbohydrates in drink. That study night was terminated, as per protocol. Another six hypoglycaemic events with glucose levels below 3.0 mmol/L occurred, five in the eating in scenario and one in the eating out scenario. Two were associated with symptoms and together with another event at 3.0 mmol/L, which also occurred before 0100, were treated with 15-18 g oral carbohydrates in drink. The remaining four events were not associated with symptoms; all but one event occurred after 0100 and were not treated. When data collected after midnight were combined (939 readings), 2 (<1%) plasma glucose values were below 2.5 mmol/L, 10 (1%) were below 3.0 mmol/L, 41 (4%) were below 3.3 mmol/L, and 72 (8%) were below 3.5 mmol/L.

No hyperglycaemic events (glucose concentrations above 16.7 mmol/l) occurred during either treatment. After midnight, all plasma glucose concentrations during closed loop delivery were below 12 mmol/L and during pump therapy were below 13.5 mmol/L.

Pooled analysis of studies

When data from both studies were combined, closed loop delivery increased the time when plasma glucose was in target by 22% (3-37%), P<0.001. Closed loop delivery reduced overnight time spent below target by 3% (0-20%), P=0.04. Closed loop delivery significantly reduced the variability in plasma glucose levels and reduced time when the levels were above target (table 5). Closed loop delivery consistently outperformed conventional insulin pump therapy at both high and low glucose values (fig 5).

Figure5

Fig 5 Distribution of plasma glucose values during closed loop insulin delivery and conventional insulin pump therapy (continuous subcutaneous insulin infusion) combining data collected from midnight until end of the eating in scenario and the eating out scenario. Percentages represent total time plasma glucose level was below, at, and above target from midnight until end of closed loop delivery

Table 5

 Comparisons between closed loop delivery of insulin and conventional insulin pump therapy based on plasma glucose concentrations after pooling results for two meal scenarios: eating in and eating out. Data are medians (interquartile ranges) unless stated otherwise

View this table:

Discussion

Overnight closed loop delivery of insulin improved glucose control and reduced exposure to low plasma glucose levels in adults with type 1 diabetes, even after a large carbohydrate meal accompanied by alcohol. The time plasma glucose levels were at or below 3.9 mmol/L was halved, and after midnight we did not observe plasma glucose values below 3.0 mmol/L. The average dose of insulin delivered during closed loop and conventional pump therapy was comparable, illustrating the key role of glucose responsive insulin delivery to achieve target glucose levels. The large evening meal and alcohol intake provided a greater challenge to closed loop delivery than did a medium sized evening meal, reflected by a slightly reduced but still superior performance over insulin pump therapy.

Strengths and limitations of the study

An important feature of our closed loop algorithm is avoidance of hypoglycaemia, which if replicated in the home setting would revolutionise patient safety. The algorithm suspends insulin delivery when the sensor detects a low or rapid fall in glucose levels and does so without increased risk of hyperglycaemia.24 Initiation of the algorithm is straightforward, requiring users to input their weight, total daily insulin dose, and basal infusion rates. The system updates advice on insulin dose using a robust, computationally efficient approach suitable for use with insulin pumps and hand held devices to facilitate practical use. It can discriminate between slowly or rapidly absorbed meals, thus coping with considerable variability in gut absorption. Similarly, variability in insulin absorption is assessed through model based analysis of insulin delivery and the effects on sensor glucose levels. These adaptive capabilities set our algorithm apart from most existing non-adaptive approaches.25 26 27

Additional strengths of our study are that we used the closed loop system in circumstances near to those in real life, without artificially improving performance. Firstly, we used a commercially available continuous glucose monitor calibrated according to manufacturer’s instructions. Secondly, only one sensor was inserted and it was not recalibrated in the event of suboptimal accuracy. Thirdly, the advice of the algorithm was always adhered to. Our studies assessed closed loop delivery of insulin in preparation for home testing, when recalibration of a sensor is impractical (particularly overnight), and multiple sensors, although improving accuracy, are inconvenient and likely to affect users’ satisfaction.

Limitations are that participants were compliant and predominantly of white ethnicity, hence applicability to less compliant people and those with poor glycaemic control is unknown. Also, we did not adjust for multiple statistical comparisons in secondary outcomes; hence these should be considered hypothesis generating rather than conclusive. Another limitation is that the research nurse manually entered sensor glucose values into the algorithm and manually adjusted the insulin pump. Presumably the lack of automation did not to affect the performance of the closed loop system but will need to be dealt with before proceeding to home testing. The evening meals were standardised for carbohydrate but not for fat or protein content, but given that identical meals were eaten on both study occasions, this should not affect the results.

Comparison with other studies

Our earlier studies described the benefits of overnight closed loop delivery of insulin in children and adolescents after a large evening meal and moderate exercise.18 They documented a 20% increase in the time that plasma glucose levels were in target and a halving of the time spent hypoglycaemic with the same closed loop algorithm.18 The present study provides further evidence that overnight closed loop delivery can operate safely, effectively, and consistently across different age groups, insulin sensitivities, and lifestyle conditions, including large evening meals, evening exercise, and alcohol intake.

After the evening intake of alcohol, we expected late morning hypoglycaemia during conventional insulin pump therapy, as observed in a previous study28 at similar ethanol concentrations. This may be attributed to differences in protocol as we opted to omit breakfast and its meal related bolus to mimic sleeping until lunchtime (not uncommon in young adults). Another difference was that in our study alcohol was consumed within two hours of the evening meal compared with three hours later in the other study. Nevertheless, our data confirm that alcohol ingestion is unlikely to interfere with closed loop delivery of insulin.

Recently, several studies of overnight and day and night closed loop delivery have been reported.25 26 27 29 None had a randomised study design and most evaluated algorithm performance rather than its clinical application. Further studies are required to quantify the barriers and benefits of novel approaches, including fully automated insulin delivery, use of glucagon,27 29 and intraperitoneal insulin delivery.30 Testing in the home setting and over several nights is now needed to determine the effectiveness on longer term glycaemic control.

Conclusions and policy implications

Introduction of closed loop systems for the delivery of insulin into clinical practice and clinical decision making may gradually tackle important clinical problems, such as reduction in the duration or severity of hypoglycaemia. As continuous glucose monitors and control algorithms are perfected, devices miniaturised, and performance limiting aspects identified, it is anticipated that early closed loop approaches such as overnight closed loop delivery will be followed by more advanced solutions for control of glucose during both the day and at night. Closed loop systems may serve as a bridge until type 1 diabetes is cured by, for example, stem cell therapy or islet transplantation.

In conclusion, closed loop delivery of insulin can significantly improve the control of glucose levels overnight and reduce the risk of nocturnal hypoglycaemia in adults with type 1 diabetes. The closed loop system has the potential to improve safety and efficacy of insulin delivery and may in future allow more flexible lifestyles in conjunction with improved glycaemic control for people with type 1 diabetes.

What is already known on this topic

  • Combined continuous glucose monitoring and insulin pump therapy improves glucose control in adults with type 1 diabetes, but requires ongoing patient education and motivation

  • Closed loop systems combine three technologies, linking real time insulin delivery with continuous glucose monitoring via a control algorithm, to replace or assist clinical decision making

  • Overnight closed loop insulin delivery reduces glucose excursions in children and adolescents after large meals and exercise, but efficacy in adults is unknown

What this study adds

  • In these two small crossover trials overnight closed loop delivery of insulin (artificial pancreas) in adults with type 1 diabetes improved glucose control and reduced exposure to low plasma glucose levels, with no overnight values below 3.0 mmol/L

  • Overnight closed loop delivery can operate safely, effectively, and consistently across different age groups, insulin sensitivities, and lifestyle conditions, including large evening meals and alcohol intake

Notes

Cite this as: BMJ 2011;342:d1855

Footnotes

  • We thank Josephine Hayes for administrative support, Angie Watts for laboratory support, Joanna Weston for dietary support, and the staff at Addenbrooke’s Wellcome Trust Clinical Research Facility for their help. Steve Luzio and colleagues measured plasma insulin.

  • Contributors: RH coordinated the studies. MLE, RH, KK, SAA, DBD, SRH, HRM, and MEW co-designed the studies. JH and KK were responsible for screening and enrolment of participants. KK and JH requested informed consent from the participants. JH, KK, JMA, and DE provided patient care, collected the clinical and laboratory data, and contributed to biochemical analysis. MEW carried out randomisation. MEW, DX, MN, RH, KK, and CK carried out or supported the data analysis, including the statistical analyses. RH designed and implemented the glucose controller. RH, KK, MLE, SAA, SRH, HRM, DBD, and MEW contributed to the interpretation of the results and the writing and critical review of the report. RH had full access to all of the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

  • Funding: This study was supported by Diabetes UK (BDA07/0003549), Juvenile Diabetes Research Foundation (22-2006-1113, 22-2007-1801, 22-2009-801, and 22-2009-802), Medical Research Council Centre for Obesity and Related metabolic Diseases, and National Institute for Health Research Cambridge Biomedical Research Centre. Smiths Medical supplied the study pumps and Abbott Diabetes Care supplied the Freestyle Navigator devices and sensors for the eating in scenario. Abbott Diabetes Care read the manuscript before submission. No sponsor had any role in the study design, data collection, data analysis, data interpretation, or writing of the report.

  • Competing interests: All authors have completed the Unified Competing Interest form at www.icmje.org/coi_disclosure.pdf (available on request from the corresponding author) and declare: in the previous 3 years RH has received speaker honorariums from Minimed Medtronic, Lifescan, and Novo Nordisk, served on advisory panel for Animas and Minimed Medtronic, received licence fees from BBraun and Beckton Dickinson, and served as a consultant to Beckton Dickinson, BBraun, and Profil; CK has served as a consultant to Medtronic International Trading Sàrl and Diabetes Technology Management; SAA has served on advisory boards for Medtronic and Johnson and Johnson and has received speaker honorariums for Medtronic, Animas, and Roche; SRH has received speaker honorariums from NovoNordisk, Eli Lilly, Sanofi-Aventis, and Lifescan and served on advisory panels for NovoNordisk and Eli Lily; MLE has received speaker honorariums from Eli Lilly and served on advisory panels for Medtronic, Sanofi-Aventis, and Cellnovo; HRM has received speaker honorariums from Minimed Medtronic; MEW has received license fees from Becton Dickinson and has served as a consultant to Beckton Dickinson; KK, JH, JMA, DE, DX, and MN declare no competing financial interests. RH, MEW, and DBD report patent applications in the area of closed loop delivery of insulin.

  • Ethical approval: This study was approved by Cambridge research ethics committee.

  • Data sharing: The supplemental appendix is available from the corresponding author at rh347{at}cam.ac.uk.

This is an open-access article distributed under the terms of the Creative Commons Attribution Non-commercial License, which permits use, distribution, and reproduction in any medium, provided the original work is properly cited, the use is non commercial and is otherwise in compliance with the license. See: http://creativecommons.org/licenses/by-nc/2.0/ and http://creativecommons.org/licenses/by-nc/2.0/legalcode.