Home blood sampling for plasma glucose assay in control of diabetes.

Br Med J 1978; 2 doi: https://doi.org/10.1136/bmj.2.6137.596 (Published 26 August 1978) Cite this as: Br Med J 1978;2:596
  1. S Howe-Davies,
  2. R R Holman,
  3. M Phillips,
  4. R C Turner

    Abstract

    Estimation of plasma glucose in home blood samples is needed to improve diabetic control. Sufficiently precise measurements on capillary blood were obtained by (a) storing Reflotest glucose-oxidase strips in a desiccant container before reading and (b) collecting blood samples into a simple vacuum bottle containing potassium fluoride (assay of sodium content indicating volume of plasma collected). The precision of the methods (+/- 1 SD) was +/-0.35 mmol/1 (+/-6.3 mg/100 ml). Clinical reliability was assessed by measuring the basal plasma glucose concentration at home on different mornings in patients with maturity-onset diabetes, the day-to-day variation (+/- 1 SD) being +/-0.73 and +/-0.92 mmol/1 (+/-13.2 and +/-16.6 mg/100 ml) respectively. The mean basal plasma glucose concentration in all 84 patients with maturity-onset diabetes from three general practices was 8 mmol/1 (144 mg/100 ml), 44 of the values exceeding 6 mmol/1 (108 mg/100 ml). Improving control by monitoring the basal plasma glucose concentration in maturity-onset diabetes might help to prevent diabetic complications.