Confirmed over-infusion: Damaged syringe pump
1. Brief Description of the Incident
-
An infusion pump was set to deliver at 10ml/hr but was reported
to deliver at 15ml/hr.
-
The over infusion was detected because of the routine practice,
by nursing staff in the ward, of checking volume infused from the syringe
barrel.
2. Device Details
-
A syringe pump whose rate is set in terms of a volumetric
flow rate and which detects the syringe size in order to convert flow rate
from the selected volume flow rate (ml/hr) to plunger movement in mm/hr.
Most syringe pumps operate in this way.
3. Investigation
-
Workshop investigations confirmed the over-infusion; at 10ml/hr
the average flow rate after an hour was 15.3ml/hr – an error of 53%.
-
The pump was tested with the BD Plastipak 50/60 ml syringes
used in the ward.
-
Closer examination then revealed that the syringe size indicator
(wrongly) recorded that a 30ml syringe was being used.
-
The pump does not require the user to verify the syringe
size when the pump is used.
-
Internal examination showed that the syringe size indicator
was damaged – the photo shows on the left a new part and, on the right,
the damaged part.
-
The "syringe size sensor arm" rests on the syringe barrel
rotating the "sensor arm flag" whose position is detected by opto-sensors
to indicate the syringe size.
-
Because of the damage (hidden within the pump), the flag
did not rotate in proportion to the syringe size and hence the 50/60ml
syringe was detected as a 30ml syringe.
-
The ratio of the cross-sectional area of the syringes (50/60ml
syringe is 5.54 cm2; 30ml syringe is 3.67 cm2), that
is 1:1.51, accounts for the ~50% higher flow rate observed.
4. Possible causes
-
Failure to note that the syringe size indicated by the pump
was not the same as the syringe being used.
-
Damaged syringe size indicator.
-
Cause of damage unknown. Damage could have been caused by
storing pumps on top of each other, or if the pump had fallen or been knocked
against another object with the force taken by the sensor.
-
A second pump was reported to the workshop from the same
area soon after this with a completely broken syringe size indicator. A
spot check revealed a further device with a sticking sensor.
5. Lessons to be learnt and remedial action to be taken
-
Report damage to medical devices. The cause
of the damage is unknown. It is vital that staff report damage to medical
devices. External appearance alone did not indicate any damage to the device.
If a device is suspected of being damaged it must be thoroughly checked
prior to its return to use.
-
Check the syringe size indicator prior to starting the pump. Know
how your syringe pump displays the syringe size and check that it correctly
detects the syringe size prior to setting up the pump.
-
Staff training to ensure that staff understand:
-
Why detection of the wrong syringe size caused the pump to over-infuse.
-
Why it is important to check the syringe size.
-
Why it is important to report damage to medical devices.
-
Why it is important to check the volume infused from the fluid container
as well as totaliser.
6. Positive actions taken by staff
-
Checking volume infused from the barrel of the syringe showing
what volume was actually delivered.
-
This ensured that staff detected that there was a fault with
the pump and minimised the duration of the over-infusion.