Alain Nordmann Medical Outpatient Division, Department
of Internal Medicine, University Hospital Basel, 4031 Basel,
Switzerland
Correspondence to: E
Battegay battegay{at}ubaclu.unibas.ch
Measuring blood pressure at home is recommended to
distinguish sustained hypertension from white coat hypertension and to monitor treatment.
1 2
Blood pressure measured at home is a better predictor of 24 hour ambulatory blood pressure monitoring, target organ damage, and cardiovascular mortality than that measured by
a doctor.3-5 We aimed to assess the accuracy of patients
measuring blood pressure at home, to identify predictors of poor
accuracy, and to determine if poor accuracy impairs assessment of hypertension.
We asked 54 consecutive patients with hypertension or
suspected hypertension (aged 30-83 years), who had been referred for 24 hour ambulatory blood pressure monitoring, to measure their blood
pressure at home twice daily between 0600 and 1000 for 30 days, and to
record the time and readings. All agreed to participate. A nurse
instructed each patient in the use of fully automated oscillometric
blood pressure monitors with integrated memory devices (Omron-IC,
Advance AG, Switzerland). To comply with the hospital's ethics
committee, we informed patients (only cursorily) about the device's
memory capacity. They were unaware that we would compare the device's
stored measurements with their reported measurements. We classified
measurements as correct if the timing of measurements (within 15 minutes) and single self reported values were identical to those
recorded by the device, or if the average of multiple measurements
differed by Overall, 2915 (89.9%) of 3240 requested measurements were
performed of which 2121 (72.8%) were reported correctly. Thirty four
patients (63%, 95% confidence interval 49% to 76%) reported 80%-100% of measurements correctly, 20 (37%) reported <80%
correctly, and 12 (22%) reported
![]()
Subjects, methods, and results
Top
Subjects, methods, and results
Comment
References
3 mm Hg for systolic or diastolic pressure.
50% correctly. The table lists the
types of mistakes. Differences of >5 mm Hg between means of reported versus recorded systolic (range 5-28 mm Hg) or diastolic (5-11 mm Hg)
home blood pressure values occurred in only eight patients (15%, 7%
to 27%), and there was no preference for inadequately reporting lower
or higher values.
Using multiple regression analysis we identified low educational level
as the only independent predictor of poor reporting accuracy (<80% of
measurements correct, P=0.004). Patients who had had 8 or less school
years had a relative risk of 3.39 (
2 test, 1.54 to 7.46)
for reporting
80% of measurements correctly compared with patients
of a higher educational level.
| |
Comment |
|---|
|
|
|---|
Our study suggests that reporting accuracy of blood pressure measurements taken at home is acceptable in most patients, but that some patients of low educational level may have poor reporting accuracy, which may affect the assessment of blood pressure.
There are some potential limitations with generalising our study. Although the number of participants was small, we suggest that most patients accurately measure blood pressure at home for two reasons. Firstly, we found only eight patients (15%, 7% to 27%) with poor reporting accuracy and, secondly, we rarely identified differences of >5 mm Hg in mean systolic or diastolic blood pressure between reported and recorded measurements. Participants were informed about the capability of the device to record measurements and this may have influenced reporting accuracy. Participants were not aware of our aim to compare their measurements with those recorded by the device, therefore a substantially biased reporting accuracy seems unlikely.
In summary, it is possible but unlikely that poor reporting
accuracy of blood pressure measurements taken at home affects the
assessment of hypertension. In less educated patients ambulatory blood
pressure monitoring may be preferable to self reported measurements.
| |
Acknowledgments |
|---|
We thank Mrs Verena Brenneisen for technical assistance and Advance AG, Switzerland for supplying the Omron-IC monitors.
Contributors: AN collected and analysed the data and wrote the paper with EB. BF participated in collecting and analysing the data. TW designed the study together with BM and EB. BM and EB initiated and supervised the research project and participated in the design and writing; they are the guarantors.
| |
Footnotes |
|---|
Funding: EB was supported by a Swiss Clinicians Opting for Research grant (32-31948.91) from the Swiss National Science Foundation.
Competing interests: None declared.
| |
References |
|---|
|
|
|---|
| 1. |
The Sixth Report of the Joint National Committee on prevention, detection, evaluation, and treatment of high blood pressure.
Arch Intern Med
1997;
157:
2413-2446 |
| 2. |
American College of Physicians.
Automated ambulatory blood pressure and self-measured blood pressure monitoring devices. Their role in the diagnosis and management of hypertension.
Ann Intern Med
1993;
118:
889-892 |
| 3. | Campbell NRC, Bass M, Chockalingam A, Lebel M, Milkovich L, for the Canadian Coalition for High Blood Pressure Prevention and Control. Self measurement of blood pressure: benefits, risks and interpretation of readings. Can J Cardiol 1995; suppl H: 18-22. |
| 4. |
Julius S, Meija A, Jones K, Krause L, Schork N, van de Ven C, et al.
"White coat" versus "sustained" borderline hypertension in Tecumseh, Michigan.
Hypertension
1990;
16:
617-623 |
| 5. | Okubo T, Imai Y, Tsuji I, Nagai K, Kato J, Kikuchi N, et al. Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population-based observation in Ohasama, Japan. J Hypertens 1998; 16: 971-975[Medline]. |
(Accepted 12 March 1999)
Read all Rapid Responses